Inhalt des Dokuments
Forward modeling of transcranial Direct Current Stimulation in Finite Element Method studies for subject specific manic depression disorder therapy
Background:
The dose of transcranial direct
current Stimulation (tDCS) is defined by electrode montage and
current, while the resulting brain current flow is more complex and
varies across individuais. The left dorsolateral prefrontal cortex
(1DLPFC) is a common target in neuropsychology and neuropsychiatry
applications, with varied approaches used to position electrodes on
subjects. TDCS is generally well tolerated; however, skin Sensation
and burns caused by poor electrode assembly design have been
reported.
Objective:
To predict brain current flow intensity and
distribution using four conventional bicephalic frontal rectangular
sponge-pad l x l electrode montages and novel HD-tDCS 3x1 and 4 xl
disc electrode ring montages that nominally targeted the 1DLPFC.
Furthermore, to investigate the impact of electrode assembly and skin
micro-architecture on dermal current density (CD) clustering. Methods:
Three MRI-scans were segmented in seven tissue compartments each in
order to assign isotropic conductivities. Four electrode positioning
methods for 5 x 5 cm2 rectangular sponge-pad electrodes over
F3-F4were evaluated. Additionally, an EEG-tDCS
cap layout containing HD electrodes was included to select electrode
ring montages for Fß targeting. Also, a scalp model was derived to
investigate dermal CD distribution. Results: The induced peak electric
field in 1DLPFC decreases with an increase in overall head volume and
by enlarging the electrode distance on the skin. Additionally, an
electrode displacement towards thinner skull structures leads to the
augmentation of cortically injected current. HD-tDCS provided enhanced
spatial focality in combination with decreased Stimulation intensity.
Electrode assembly designs with a housing that granted an extensive
contact area and a large gel volume of a low conductivity resulted in
decreased CD concentrations.
Conclusion:
Cortical electric field distribution fluctuations,
for a given dose, are a function of inter-individual differences and
pose the need for subject specific tDCS therapy. Novel, reproducible,
ease-to-use electrode positioning methods were introduced to optimize
1DLPFC targeting. HD electrode ring montages for a novel closed loop
EEG-tDCS device were found and electrode assembly design parameters
were advanced.
/Bild1.jpg